| Probe |
| Connectors |
12-Way Lemo 2B (IP68) (Absolute, Bridge and Reflection) and Connection Lemo 00 (IP68) (for single element absolute probes). Simultaneous probe operation possible using Lemo 12-Way and Lemo 00. |
| Rotary Drive |
600-3000 rpm – ETher Mercury Drive (ADR002), Hocking 33A100, Rohmann MR3, SR1 & SR2 Drive (special adpater needed) |
| Conductivity |
Option becomes active with use of an AeroCheck Conductivity probe and cable (see end of spec table) |
| Frequency |
| Single/Dual |
Single 10Hz – 20MHz with range variable resolution. Dual 10Hz – 20MHz |
| Gain |
| Overall |
-18 to + 104 dB, 0.1, 1 and 6dB steps (104dB maximum) + Mix Gain (-18 to +18dB on Output) |
| Input |
0dB or 12dB |
| Drive |
– 6dB to 10dB in 1dB steps (0dB reference 1mW into 50 ohm) |
| Max X/Y Ratio |
+/-100.0dB |
| Phase |
| Range |
0.0-359.9°, 0.1° steps |
| Auto Phase |
Allows phase angle to be automatically set to a pre-set angle |
| Filters |
| Normal High Pass |
DC to 2kHz or Low Pass Filter, which ever is the lower in 1 Hz steps. Plus variable adaptive balance drift compensation 0.01 – 0.5 Hz (6 steps) |
| Normal Low Pass |
1Hz to 2kHz or a quarter of the lowest test frequency, which ever is lower in 1 Hz steps |
| Balance |
| Manual |
14 internal balance loads; 2.2µH, 5.0µH, 6.0µH, 6.5µH, 7.0µH, 7.5µH, 8.2µH, 12µH, 15µH, 18µH, 22µH, 30µH, 47µH, 82µH |
| Automatic |
Optimised balance load selection |
| Alarms |
| Box & Sector |
Both Alarm types are fully configurable, Freeze, Tone or Visual |
| Output |
Open collector transistor (50v dc at 10mA max) available on 12-way Lemo |
| Display |
| Type |
145mm (5.7”), 18 bit Colour, daylight readable |
| Viewable Area |
115.2mm (4.53”) (Horizontal) x 86.4mm (3.4”) (Vertical) |
| Resolution |
640 x 480 pixels |
| Colour Schemes |
User configurable Dark, Bright and Black & White |
| Configurable Screen |
Full Screen, Single, Dual Spot or Dual Pane with variable size and location and function e.g. XY, Time-base, Waterfall and Meter. |
| Display Modes |
Full Screen, Single, Dual Spot or Dual Pane with variable size and location and function e.g. XY, Time-base, Waterfall and Meter. Spot, Time base (0.1-20 seconds x 1-200 sweeps and up to 55 seconds), Waterfall and Meter with peak hold and % readout |
| Graticules |
None, Grid (4 sizes 5, 10, 15 and 20% FSH), Polar (4 sizes 5, 10, 15 and 20% FSH) |
| Offset |
Spot Position: Y =-50 to +50, X =-65 to +65% |
| Digital Spot Display |
in X, Y or R,θ |
| Setting |
Display/Edit of all settings in Legacy Format |
| Removable Storage Data |
| Setup Storage |
micro SD up to 32GB, holding over 10,000 settings |
| Stored Screen Shots |
micro SD up to 32GB, holding over 10,000 screen shots |
| Shots |
Comprehensive Record, Replay and Storage |
| Outputs |
| Record Replay |
Real-time recording of trace data and Replay on instruments and desktop PC up to 164 seconds |
| PC Connectivity |
USB (Full PC remote control plus Real Time data) |
| Digital Volt Free Alarm |
On Lemo 12-way Open collector transistor (36v dc at 10mA max) |
| VGA |
Full 15 way VGA output |
| Languages |
|
English, French, Spanish, Italian, Portuguese, Russian, Japanese, Chinese, Turkish, Czech, Norwegian. |
| Verification Level |
|
The system includes on delivery a 2 year validity Verification Level 2 detailed functional Check and calibration, as per ISO 15548-1:2013. |
| Power On Self Test |
|
A “self test” on start-up is performed of external ram, accelerometer, Micro SD card, LCD screen buffer. |
| Power |
| Battery |
Internal 7.2V nominal @ 3100mAh = 22.32 watt.hr |
| Running Time |
Over 7 hours with a Weld Probe at 100kHz and 50% backlight |
| Charging Time |
2.5 hrs. charge time, simultaneous charge and operation |
| External |
100-240v 50-60Hz 30 Watts |
| Connector |
Lemo OS Hermaphroditic keying, half-moon insert (IP68) |
| Physical |
| Weight |
1.15 kg (2.54 lbs) |
| Size (w x h x d) |
222.2mm x 152.2mm x 47.4mm (LxHxW) (8.75” x 6.0” x 1.87”) |
| Material |
Main Body: PC-ABS a blend of the two polymers – Polycarbonate (PC) and Acrylonitrile Butadiene Styrene (ABS). Over-moulded Material: TPE Red Rubber, Thermoplastic Elastomer (TPE). |
| Operating Temp. |
-20 to +60°C (-4 to 140 °F) |
| Storage Temp. |
Storage for up to 12 months -20 to +35°C (-4 to 95°F) Nominal +20°C (68°F) |
| IP Rating |
Designed to meet requirements of IP64 |
| Advanced Features |
| Guides |
Create and display a slide show containing instructions, tutorials and procedures using Microsoft PowerPoint. |
| Attachments |
Screenshots and Data Recordings are saved in a folder with the name of the Settings. |
| Loop |
Capture a live repetitive signal and then optimise instrument settings (Phase, Gain, Filters) to simplify optimising the parameters |
| Trace |
Allows a calibration reference signal to be stored on the screen, which can then be compared with a live signal. |
| Data Output |
Real-time, post processed over USB at 8kHz overall for all 3 data pairs (X, Y and Mix) with DLL for embedding functionality into software. |
| Conductivity Features |
| Frequency/Resolution |
60kHz – 3 decimal points max Auto Resolution Mode AutoS = legacy instrument, Auto = SigmaCheck |
| Accuracy |
0.5%-10% IACS better than +/-0.05% IACS; 10%-25% IACS better than +/-0.25% IACS; 25%-60% IACS better than +/-0.5% IACS; 60%-110% IACS better than +/-1% IACS; Lift Off corrected to 1.0mm; No temperature compensation; All Errors at 90% Confidence Level |
| Resolution |
3 decimal points max |
| Auto Resolution Mode |
AutoS = Legacy Instrument, Auto = SigmaCheck |
| Optional |
|
| Thumbwheel |
Number of Detent 12 Material Polyamide, polycarbonate. |